BLP-10 Glass bottle internal pressure tester Introduction to the instrument: The internal pressure resistance tester for glass containers is developed and produced according to the national standard GB/T4546-2008 internal pressure test method for glass containers. Comply with the international standard ISO7458-2004. Detection method: Method A: Apply a uniform internal pressure test within a predetermined period of time, and the bottle is qualified if it is not broken for 60 seconds. And incremental experiments that gradually increase the pressure. Method B: Increase the internal pressure according to the predetermined constant speed test, Pr=1.38P60+0.1783, increase the pressure to Pr, no pressure retention, pressure immediately relief, the bottle does not break is qualified! When the sample bottle is broken midway, the peak of the maximum destructive force of the bottle is recorded. Usage: Mainly used for glass bottles, glass infusion bottles. Technical parameters: 1.The machine can be "pass", "incremental", "destructive" test requirements; 2.Real-time display of pressure holding value, damage peak value and the corresponding P60 value, test times, Chinese character printout; 3.Sensor range: 0-6Mpa, safety overload 7.2Mpa; 4.Comprehensive error: <0.5%FS5. Display accuracy: 0.001Mpa; 6.Working temperature: 5 -- 40ºC; 7.Linear rate supercharging: 0... 58±0.1Mpa/s(4±1bar/s); 8.Pressure medium: water; 9.Power supply: 220V50HZ rated power 0.6KW; 10.Display output: touch screen testing system display measured value, set value, P60 value and times; Usage characteristics: 1. High measurement accuracy, direct detection of bottle pressure, real-time display, fast working frequency, high stability! Large range, the maximum set test pressure is 6.00Mpa; 2. The test speed is fast. The test time for each bottle should not exceed 8 seconds; 3. Glass and water separation, easy to work, water can be directly discharged into the sewer pipe;