Customization: | Available |
---|---|
After-sales Service: | Online Service |
Warranty: | 1 Year |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
1. | Single push rod dilatometer for measurement of length changes (expansion and/or shrinkage) along with operating and data acquisition software, hardware and calibration standards. | |
2. | Temperature range | Room temperature ~ 1600ºC |
3. | Sample diameter | 6~10mm |
4. | Sample length | 50mm(max) or 25mm |
5. | Atmosphere | Argon |
6. | Measuring range and error | 5000 um ± 0.1% |
7. | Heating element | Molybdenum wire up to 1700ºC |
8. | Sample holder and probe rod | Alumina |
9. | Dilatometer orientation | Horizontal |
10. | Expansion/shrinkage value measuring resolution | 20nm |
11. | Thermocouple | Type B |
12 | Water cooling | Closed loop water cooling unit |
13 | Heating rate | 1 ~ 30 oC /minute, accuracy±1 oC |
14 | Calibration standards | Alumina standard sample |
15 | Temperature control | Experimental process controls parameters, including PID parameters and the way of communication. Arbitrary adjust, set, lock are realized by intelligent instrument in the whole process. Modifying the parameters is achieved by the host computer (the computer). |
16 | Data Acquisition | Through onboard computer with printer, software can work in platform of Windows 10/7/XP. |
17 | Power supply | 220V, 50HZ |